INTELLIDOX DOCKING MODULE USER MANUAL || DETECTOR OPERATIONS
HONEYWELL
PAGE 54 OF 119
Calibration Guidelines
1. Perform compliance tests only in a normal environment that is 20.9% v/v O
2
and free
of hazardous gas. Do not operate the module in a hazardous area. Failure to adhere
to this guideline can result in possible personal injury and/or property damage.
2. The use of calibration gas cylinders other than those specified by Honeywell may
result in unsafe calibration or possible non-recoverable failure of the equipment, and
will invalidate the warranty.
3. Do not use gas cylinders past their expiration date.
4. A demand flow regulator must be used with all gas cylinder connections. The module
inlet should not be pressurized.
5. Do not calibrate the detector immediately after charging is complete.
6. Calibrate the sensor if ambient gas readings vary during startup.
7. Calibrate a new sensor before use. Allow the sensor to stabilize before starting
calibration.
a) Used sensor: wait 60 seconds
b) New sensor: wait 5 minutes
8. When calibrating the same detector multiple times, wait 10 minutes between
calibrations to allow the sensor to stabilize.
9. If a certified calibration is required, contact Honeywell or an authorized distributor.