INTELLIDOX DOCKING MODULE USER MANUAL || ASSEMBLE MODULES
HONEYWELL
PAGE 15 OF 119
Work Plan: Assemble Groups of up to Five Connected Modules
Up to five modules may be connected to form a group. Connected modules share the power
supply, exhaust tubing, inlet filter, connected test gas cylinders and network connection. Follow
these steps to assemble and prepare groups of up to five connected modules for first use:
1. Place the modules and component parts on a clean, dry work surface.
2. Verify that the appropriate inlet keys are inserted. Replace the inlet keys if necessary. For
more information, see
3. Assemble the stand or mount the module on a wall or DIN rail if necessary. For more
information, see:
•
•
•
4. Connect up to 5 modules. For more information, see:
•
•
Mounting Connected Modules on a Wall
5. Prepare the connected modules for use. For more information, see
a) Attach the end plate to the last module in the group. See
b) Connect an inlet filter to the purge inlet on the first module in the group. For more
information, see
c) Insert an inlet plug into each unused inlet on the first module in the group. For
more information, see
d) Connect the exhaust tubing to the first module in the group. For more
information, see
e) Connect the power to the first module in the group.
f) Connect to a network. See
Connect the Module to a Network
6. Connect the calibration gas cylinder or cylinders, and then configure the gas inlets. For
more information, see