INTELLIDOX DOCKING MODULE USER MANUAL || PREPARE MODULES FOR USE
HONEYWELL
PAGE 36 OF 119
Connecting a Calibration Gas Cylinder
1. Connect a demand flow regulator to the calibration gas cylinder, and then attached the
supplied connector with attached 3/16 inch (9.5 millimeter) I.D. tubing to the demand flow
regulator.
2. Cut 1/8 inch (3.2 millimeter) I.D. tubing to an appropriate length. Honeywell recommends
that the tubing is between 39 inches (1 meter) and 33 feet (10 meters) in length.
3. Attach one end of the 1/8 inch (3.2 millimeter) I.D. tubing to the demand flow regulator.
4. Insert a quick connect fitting into the free end of the tubing, and then connect it to Inlet 1
on the module.
5. If a multi-inlet key is inserted in an individual module or one or more connected modules in
a group, then repeat steps 1, 2, 3 and 4 as necessary until up to 4 cylinders are attached.
6. Configure the gas inlets before using the module. See