INTELLIDOX DOCKING MODULE USER MANUAL || ASSEMBLE MODULES
HONEYWELL
PAGE 23 OF 119
Mounting Modules on Parallel DIN Rails
1. Prepare the modules for mounting.
a) Disconnect modules that are connected into groups.
b) Disconnect the power cord, network cable, tubing and inlet filter assemblies.
c) Remove the end plate.
2. Attach a module to the DIN rails. Repeat as necessary until all modules attached.
a) Hold the module upright with the back panel toward the DIN rails. Place the top
DIN mount over the top of the upper DIN rail and then lower the module toward
the lower DIN rail until it snaps into place.
b) Verify that the module is firmly attached to both the upper DIN rail and the lower
DIN rail.
3. Connect the attached modules into groups of up to 5, if using. Repeat as necessary until
all groups are connected.
a) Snap the first module into position on the DIN rail, and then unlock and raise the
latch arm.
b) Snap the second module onto the DIN rail, and then slide it into position beside
the first module. Push the modules firmly together and then lower and lock the
latch arm.
c) Repeat as necessary until the group is complete.
4. Prepare the module for use. For more information, see
a) Attach the end plate. Verify that the latch arm is lowered and locked.
b) Connect the inlet filters, tubing, network connection and power cord.