INTELLIDOX DOCKING MODULE USER MANUAL || DETECTOR OPERATIONS
HONEYWELL
PAGE 60 OF 119
Charge a Detector
You may use the module to charge compatible detectors that are fitted with rechargeable
batteries. For more information on battery maintenance, refer to the appropriate detector
operator manual.
Charge only in a normal environment that is 20.9% v/v O
2
and free of hazardous gas.
Do not operate the docking module in a hazardous area. Failure to adhere to this
guideline can result in possible personal injury and/or property damage.
Charging a Detector
Battery charging is disabled during bump test and calibration procedures.
Do not store the detector in the module.
1. Insert a compatible detector into the module.
Detector identification
is displayed
on the LCD.
2. When the detector is recognized, procedures specified to occur automatically on
insertion are performed. If the detector contains a rechargeable battery, battery
charging begins.
3. After the procedures are completed, you may leave the detector in the module for
charging. If the detector is activated, the module will deactivate it after 10 minutes
of inactivity and continue charging the battery.
4. When charging is complete, remove the detector from the module.