10 Technical Data
10.1 General Information
dimensions/
housing
•
plug-in unit in 19” technology, 3 height units
(3U), 1/3 or 1/2 housing width
•
depth without plug approx. 170 mm, with plugs
approx. 220 mm
•
process interfacing at the rear, operation
panel at the front
ambient conditions
•
ambient temperature range −10 °C up to
+55 °C
•
humidity < 90 %, non-condensing
•
installation outside Ex zones 0, 1 and 2 only
IP protection class
•
IP20
storage temperature
•
−25 °C to +60 °C
weight
(fully equipped)
•
1/3 device width:
approx. 1.3 kg
•
1/2 device width:
approx. 2.1 kg
power supply
•
24 V DC with a tolerance of -15 % and +20 %
(i.e. maximum permissible range 20.4 V DC to
28.8 V DC),
power consumption up to 12 W (typically 5 W)
•
optional: 230 V AC via external power supply
•
operation in conjunction with an
uninterruptable power supply is recommended
(
6.2 Power Supply and Grounding
p. 47)
display/operation
panel
•
4.3” TFT color display with touch screen,
480×272 pixels RGB, LED background
lighting
Содержание enCore FC1