MAN-08723-001 Rev. 001 page 5 of 28
G.2
DESCRIPTIVE DIAGNOSIS SENSITIVITY AND SPECIFICITY ESTIMATES
A panel of three independent Cytopathologists adjudicated slides from all discordant (one-grade or
higher cytologic difference) descriptive diagnosis cases (639), all concordant positive cases (355)
and a random 5% subset of the 8550 negative concordant cases (428). The Cytopathologists on the
adjudication panel were board-certified, all of whom had a subspecialty certification in
Cytopathology. Their experience levels in Cytopathology ranged from 6 to 12 years. Two of the
adjudicators were from university practices and one adjudicator was from a private medical center.
The volumes for the adjudicator’s institutions ranged from 12,000 to 30,000 ThinPrep
®
Pap Tests
annually.
A consensus diagnosis was defined as agreement by at least 2 of 3 Cytopathologists. All slides sent
to the panel of Cytopathologists were not identified by site nor ordered in any fashion. When a
consensus diagnosis could not be obtained by at least 2 of 3 Cytopathologists, the full panel of
Cytopathologists reviewed each case simultaneously using a multi-headed microscope to determine
a consensus diagnosis.
The adjudicated results were used as a “gold standard” to define the following major “true”
descriptive diagnosis classifications of the Bethesda System: Negative, ASCUS, AGUS, LSIL,
HSIL, Squamous Cell Carcinoma (SQ CA) and Glandular Cell Carcinoma (GL CA). Estimates of
sensitivity and specificity together with 95% confidence intervals were calculated for the
Manual
Review
and
Imager Review
arms of the study. The differences in sensitivity and specificity between
the two arms, together with their 95% confidence intervals were also calculated. Among the random
5% subset of 8,550 cases (428 slides) that were found to be negative by both arms and adjudicated,
there were 425 “true” negative and 3 “true” ASCUS slides. A multiple imputation technique was
used to adjust the numbers of true positives and true negatives for the 8,550 negative concordant
cases based on the 5% of cases that were adjudicated
3
.
Tables 2-4
below summarize the descriptive diagnosis sensitivity and specificity estimates with 95%
confidence intervals for each of the four sites and all sites combined for “true” ASCUS+, LSIL+
and HSIL+.
Table 2: Adjudicated Review Versus Imager And Manual Reviews ASCUS+
Descriptive Diagnosis Summary.
Sensitivity is a percent of “true” ASCUS+ (combined ASCUS, AGUS, LSIL, HSIL, SQ CA and GL CA) slides
classified in either study arm as ASCUS+ and specificity is a percent of “true” Negative slides classified in either
study arm as Negative.
Sensitivity
Specificity
Site/
Number
Cases
Manual
Imager
Difference
Site/
Number
Cases
Manual
Imager
Difference
Site 1
77.2% 78.3% +1.1%
Site 1
98.7% 99.2% +0.4%
180
(70.4, 83.1)
(71.6, 84.1)
(-5.8, 8.0)
2132
(98.1, 99.1)
(98.7, 99.5)
(-0.1, 1.0)
Site 2
63.1% 77.5% +14.4%
Site 2
95.8% 96.1% +0.3%
230
(56.5, 69.3)
(71.4, 82.6)
(8.2, 20.5)
2210
(94.9, 96.6)
(95.2, 96.9)
(-0.7, 1.3)
Site 3
80.6% 94.2% +13.6%
Site 3
98.5% 98.8% +0.4%
103
(71.6, 87.7)
(87.8, 97.8)
(4.3, 22.9)
2196
(97.9, 99.0)
(98.3, 99.2)
(-0.3, 1.0)
Site 4
87.2% 84.4% -2.8%
Site 4
97.3% 97.0% -0.3%
179
(81.4, 91.7)
(78.2, 89.4)
(-10.6, 5.0)
2313
(96.6, 97.9)
(96.2, 97.7)
(-1.1, 0.5)
All
75.6% 82.0% +6.4%
All
97.6% 97.8% +0.2%
692
(72.2, 78.8)
(78.8, 84.8)
(2.6, 10.0)
8851
(97.2, 97.9)
(97.4, 98.1)
(-0.2, 0.6)
Numbers in parentheses represent 95% confidence intervals.
The results presented in
Table 2
show that for ASCUS+, the increase in sensitivity of the
Imager
Review
over the
Manual Review
was statistically significant with the lower limit of the 95%
confidence interval being 2.6% for all sites combined. The observed difference between sensitivities
for ASCUS+ varied among the sites from –2.8% with a 95% confidence interval of (– 10.6%; 5.0%)
to +14.4% with a 95% confidence interval of (8.2%; 20.5%). The difference in specificity results
between the
Imager Review
and the
Manual Review
was not statistically significant with a 95%
confidence interval of -0.2% to +0.6%. The observed differences between specificities varied among
the sites from –0.3% to +0.4%.
Содержание ThinPrep
Страница 1: ...Operator s Manual ThinPrep Imager Duo Imaging Station...
Страница 4: ...The ThinPrep Imaging System The ThinPrep Imaging System...
Страница 5: ...Operation Summary and Clinical Information The ThinPrep Imaging System...
Страница 33: ...Table of Contents Table of Contents...
Страница 37: ...iv Imager Duo Operator s Manual This page intentionally left blank...
Страница 38: ...1 Introduction 1 Introduction...
Страница 41: ...INTRODUCTION 1 Imager Duo Operator s Manual 1 3 ThinPrep Imager Duo System Laboratory Flow Figure 1 2 Lab Flow...
Страница 60: ...1 INTRODUCTION 1 22 Imager Duo Operator s Manual This page intentionally left blank...
Страница 61: ...2 Installation 2 Installation...
Страница 66: ...3 Operation 3 Operation...
Страница 99: ...4 Software Menu 4 Software Menu...
Страница 153: ...4 OPERATION OF THE SOFTWARE MENU 4 54 Imager Duo Operator s Manual Figure 4 69 Rate Report Window...
Страница 171: ...4 OPERATION OF THE SOFTWARE MENU 4 72 Imager Duo Operator s Manual This page intentionally left blank...
Страница 172: ...5 Maintenance 5 Maintenance...
Страница 180: ...5 IMAGER DUO MAINTENANCE 5 8 Imager Duo Operator s Manual This page intentionally left blank...
Страница 181: ...6 Troubleshooting 6 Troubleshooting...
Страница 201: ...6 TROUBLESHOOTING 6 20 Imager Duo Operator s Manual This page intentionally left blank...
Страница 202: ...7 Defi nitions and Abbreviations 7 Defi nitions and Abbreviations...
Страница 206: ...7 DEFINITIONS AND ABBREVIATIONS 7 4 Imager Duo Operator s Manual This page intentionally left blank...
Страница 207: ...8 Service Information 8 Service Information...
Страница 209: ...8 SERVICE INFORMATION 8 2 Imager Duo Operator s Manual This page intentionally left blank...
Страница 210: ...9 Ordering Information 9 Ordering Information...
Страница 213: ...10 Remote Site Server User s Guide 10 Remote Site Server User s Guide...
Страница 222: ...Index Index...
Страница 229: ...Operator s Manual ThinPrep Imager Duo Imaging Station...