T
ROUBLESHOOTING
6
Imager Duo Operator’s Manual
6.17
Writable media verification failed.
After the media was written to,
the data didn't match what was
exported.
Try the function again.
Database image too large for media.
Too much data to fit on selected
media type.
If this message appears during database
backup, consult Hologic Technical Support.
Database query failed.
(when viewing Import Results)
A database query failed.
Try the function again. If the error persists, con-
tact Technical Support.
Export failed.
The export failed.
Try again with a different DVD/CD or USB,
depending on what type of media was used.
Make sure the storage media is in the Server
drive.
Export failed due to database error.
A database error prevented the
export from succeeding.
Try the function again.
Failed to add the template.
A new template could not be
added to the Export menu.
Try to add a template again. Be sure it has a
unique name.
Failed to write to the media.
The media type failed to write
properly.
Try again with a different DVD/CD or USB,
depending on what type of media was used.
Make sure the storage media is in the Server
drive.
Failed to delete the template.
The template could not be
deleted.
It may have already been deleted by another
user.
Failed to modify the template.
A template could not be modi-
fied.
The template has been deleted. It must be rec-
reated.
Failed to start the task.
The server was unable to begin
the requested import/export/
backup.
Try the function again.
Failed to update FOI table.
An error occurred during data
import.
Contact Technical Support.
Failed to update SlideImage table.
An error occurred during data
import.
Contact Technical Support.
Import cannot be performed from the
media.
The storage media in the
server's drive does not have
export data on it.
The storage media may have other data on it.
Only data exported from a ThinPrep Server will
be imported.
Import failed.
The import failed.
Try the import again.
Make sure the storage media is readable.
Table 6.3: Data Export, Import and Backup Errors
Error Text
Possible Cause
Corrective Action
Содержание ThinPrep
Страница 1: ...Operator s Manual ThinPrep Imager Duo Imaging Station...
Страница 4: ...The ThinPrep Imaging System The ThinPrep Imaging System...
Страница 5: ...Operation Summary and Clinical Information The ThinPrep Imaging System...
Страница 33: ...Table of Contents Table of Contents...
Страница 37: ...iv Imager Duo Operator s Manual This page intentionally left blank...
Страница 38: ...1 Introduction 1 Introduction...
Страница 41: ...INTRODUCTION 1 Imager Duo Operator s Manual 1 3 ThinPrep Imager Duo System Laboratory Flow Figure 1 2 Lab Flow...
Страница 60: ...1 INTRODUCTION 1 22 Imager Duo Operator s Manual This page intentionally left blank...
Страница 61: ...2 Installation 2 Installation...
Страница 66: ...3 Operation 3 Operation...
Страница 99: ...4 Software Menu 4 Software Menu...
Страница 153: ...4 OPERATION OF THE SOFTWARE MENU 4 54 Imager Duo Operator s Manual Figure 4 69 Rate Report Window...
Страница 171: ...4 OPERATION OF THE SOFTWARE MENU 4 72 Imager Duo Operator s Manual This page intentionally left blank...
Страница 172: ...5 Maintenance 5 Maintenance...
Страница 180: ...5 IMAGER DUO MAINTENANCE 5 8 Imager Duo Operator s Manual This page intentionally left blank...
Страница 181: ...6 Troubleshooting 6 Troubleshooting...
Страница 201: ...6 TROUBLESHOOTING 6 20 Imager Duo Operator s Manual This page intentionally left blank...
Страница 202: ...7 Defi nitions and Abbreviations 7 Defi nitions and Abbreviations...
Страница 206: ...7 DEFINITIONS AND ABBREVIATIONS 7 4 Imager Duo Operator s Manual This page intentionally left blank...
Страница 207: ...8 Service Information 8 Service Information...
Страница 209: ...8 SERVICE INFORMATION 8 2 Imager Duo Operator s Manual This page intentionally left blank...
Страница 210: ...9 Ordering Information 9 Ordering Information...
Страница 213: ...10 Remote Site Server User s Guide 10 Remote Site Server User s Guide...
Страница 222: ...Index Index...
Страница 229: ...Operator s Manual ThinPrep Imager Duo Imaging Station...