6 System integration
Operating manual mzr-2521 and mzr-2921
44
Technical data subject to change without prior notice!
Last update: June 2019
Remarks:
−
It is possible to adjust speed of the micro annular gear pump with the
potentiometer without the need to connect the serial interface.
−
With the analog nominal value input (connection clamps »AnIN« and »GND«)
it is possible to adjust speed of the pump with a standard signal 0-10°V. For
this purpose it is necessary to plug the jumper on the S-BL control unit from
the »AnalogPoti« to the »AnalogExtern«. The serial interface does not need to
be connected.
−
In case of an overcurrent error the green status LED on the S-BL control unit
turns red
−
The standard programs memorized in the motor control unit may be started
with the tumbler switch S1. Basic sample programs are shown in the
chapter 7. For more advanced programs please refer to the user manual for
Motion Controller MCBL3006.