12
Process without time setting
Only set a set temperature and a set shaking
frequency but set the time on continuous
00:00.
Now, the process time is displayed
continuously. It only starts to run if the tem-
perature set value is reached. A set value
change can also be made as described above
during a running process. Changes are re-
tained also after the completion of the pro-
cess and are displayed in the start menu as
current set values. For terminating the pro-
cess, push
Stop
.
Process with time setting
If a process time is set, the remaining time is
displayed in the menu. The time only starts to
run if the temperature set value is reached.
This does also apply for subsequent tempera-
ture changes. A set value change can also be
made as described above during a running
process.
The process ends without a signal or contin-
uous with a signal until it is terminated
manually. The relevant modus can be set in
the setup menu.
For changes during the process
If set value changes have been made, you can
select to save these changes and to have
them displayed in the start menu as new set
values.
Stopping the process before the expiration
of the time
If the process is stopped or changed before
the expiration of the time, you can select
whether the process really should be termi-