Interval shaking =
Interval shaking is an alternative when the probe shall not
be subject to permanent shaking. A resting time of up to 9.9 minutes can be set in
"pause". The following short shaking time of up to 9 seconds is set in "mix sec".
This is a reasonable choice for the last program running of a complete program with
time setting "endless" during off working hours.
Pause
[min]
=
Resting time of up to 9.9 minutes. If no intervals are to be set, enter
the value 0.0 here.
Duration of pause is 0.0 as standard. If no changes are to be made, select the next
field "step" by using the
–
(down) button. To change this value press
ok
; the ones
column is highlighted and can be changed by using buttons
+
and
–
. For changing
the 1/10 digit press button
>
; the digit is highlighted and can be changed with but-
tons
+
and
–
. Confirm your entry by pressing
ok
and the next field "mix sec" will be
blinking.
Mix [s]
= Short shaking time up to 9 seconds after resting time.
For "mix sec" 1 is set as standard. If no changes are to be made, select the next field
"step" by using the
–
(down) button. For changing the shaking time press
ok
and
change time by using buttons
+
and
–
. Confirm your entry by pressing
ok
and the
next field "step" will be blinking.
Entry of further program steps
If another program step shall be added to
a chosen program, press
ok
at the blin-
king field "step" and increase the number
of the last program step by 1. Further
entries as already described.
To change the program data entry
To change the chosen program, enter the
program step number to be changed.
Select the field to be changed by using
buttons
+
and
–
. Press
ok
for changing.
Changes are made as already described