9
Rotation direction: right, left
Page 2:
Contrast: Mentioned in %
Brightness: Mentioned in %
Display colour: Normal - inverse
10-point-calibration
By default the standard calibration begins
with the block number “B0”. 10 temperature
bases are underlying this calibration. The
calibration is optimized for all DITABIS stand-
ard blocks. If you are using customized blocks
you will be able to do individual calibrations
in addition. For the calibration you need a
gauged high-precision temperature measur-
ing device.
Select block number in the set up menu and
confirm with
ok
. You are now able to cali-
brate and save different individual calibra-
tions, from standard block B1 up to 4. There-
fore please select B1 to B4. Confirm again
with
ok
. Now you will get to the calibration
menu. In the first and third column you can
see the temperature base. In the second and
fourth column you can see the variable set
values.
By pushing the
+
and
–
keys the entries and
values can be changed. The selected field is
blinking; by pushing
ok
the field is selected
for changing.
Example:
You have adjusted the device for 95,0°C in the
start menu and are measuring only 85,0°C