10
with a calibrated thermometer inside your
sample. Now you have to change the temper-
ature set value in column four about 10°C.
By doing so you can calibrate every single
base individually.
Temperature set value
Navigate with
+
and
–
to the temperature
input box and confirm with
ok
. The individual
numbers can be selected with
arrow left
and
right
, their value can be changed with
+
and
-
.
The input is confirmed with
ok
and the
cursor automatically jumps to the
Start
but-
ton.
A set value change can also be made in the
same way during a running process. After
confirming the input with
ok
, the new set
value is effective immediately. For the hun-
dreds, the following values can be selected:
0 for temperatures up to +99,9°C
1 for temperatures from +100,0°C
- for temperatures below 0,0°C
Important information
For cooling systems, the minimum tempera-
ture may not be smaller than the difference
from room temperature mentioned below.
MKR 13 / TK 23: 16°C below r.t.
MKR 23: 11°C below r.t.
Mix set value
Navigate with
+
and
–
to the shaking fre-
quency input box and confirm with
ok
. The
individual numbers can be selected with
arrow left
and
right
, its value can be changed
with
+
and
–
.
The input is confirmed with
ok
and the cursor