5 - 41
Code Error
Message
Countermeasure
1390 The focus position is not within
a proper range.
Adjust the specimen stub
height to proper height.
This message appears when the inaccurate
focusing due to the specimen holder located too
low lasts for 2 minutes during HV-ON. Click the
[OK] button to close the message. Stop image
observation once and readjust the height of the
specimen holder properly.
For specimen preparation and height
adjustment, refer to 4.2.2 in the instruction
manual.
1395 The focus position is not within
the proper range.
Adjustment of the specimen
stub height is recommended.
Do you continue the
processing?
This confirmation message appears if the focus
position is not proper for this operation (causing
a large magnification error) while trying to
perform Quick Save, Save or Digital Zoom.
Click the [OK] button to continue the current
process; click the [Cancel] button to cancel the
process.
It is recommended to stop image observation
once and readjust the height of specimen holder
properly. For specimen preparation and height
adjustment, refer to 4.2.2 in the instruction
manual.
1401 Time has run out. “Image
Capture” cannot be executed.
Start up the main unit and
application again. Contact
service if the failure occurs
repeatedly.
This message appears if a time-out (50 s) has
occurred during image capturing. Image
capturing stops and the View mode returns to
the previously selected mode. If the message
appears repeatedly, contact the nearest service
representative.
1411 Time has run out. “Quick
Save” cannot be executed.
Start up the main unit and
application again. Contact
service if the failure occurs
repeatedly.
This message appears if a time-out (30 s) has
occurred during Quick Save. Quick Save stops
and the view mode returns to the previously
selected mode.
If the message appears repeatedly, contact the
nearest service representative.