7 - 76
4.
The Adjustment dialog shown in Figure 7.24.1-2 is displayed. Input the calibration value
of horizontal direction (X) and vertical direction (Y) between two points, or the angle
between two points. And then, click the [Next] button.
NOTE
: In case of inputting the angle, the value of vertical direction (Y) is calculated and
calibrated by the value of horizontal (X) and input angle.
Figure 7.24.1-2 Adjustment of X, Y Dialog
5.
The confirmation message shown in Figure 7.24.1-3 is displayed. When clicking the [OK]
button, [X, Y] calibration is executed. The left side of the [X, Y] button in the [Calibration]
block changes into an orange mark.
Figure 7.24.1-3 Confirmation Dialog
X, Y calibration ends with the above-mentioned procedure.
When selecting the [X, Y] from the [Calibration] menu or clicking the [X, Y] button on the
[Calibration] block with X, Y calibration executed, the content of X, Y calibration is
cleared and returned to the default.