13. Trouble shooting
Symptom
Possible cause
Solution
Fluid leaks at the gap
- Valves loose or damaged
- Tighten valves with the wrench or replace
between the valve
if necessary (see point 10)
Note:
After
block and the valves
sterilization (see point 9), suction/discharge
/recirculation valves need to be tightened.
Fluid is leaking
- Discharge tube is loose
- Mount the original discharge tube correctly
between the discharge
or damaged
(see point 6.2).
tube and the valve
- Wrong discharge tube
- Use only the original discharge tube.
used
Device aspirates
- Device not primed
- Prime the device (see point 7.2).
air
- Suction tube loose
- Tighten suction tube (see point 6.1).
- Suction tube or retainer
- Replace suction tube (see accessories
ring damaged
spare parts point 15).
- 48 -