current calibration factor. The default setting is
20.00 for a 20 ml device and 50.00 for a 50 ml
device. The calibration factors may be set between
18.00 and 21.99 for a 20 ml device and between
45.00 and 54.99 for a 50 ml devices.
Pressing the right key modifies the actual decimal
position flashing at the time. When the left key is
pressed the device switches to the next decimal
position.
After calculating the actual volume of the dispensed
fluid (ideally use the maximum burette volume), the
new factor can be entered directly as the customer-
specific calibration factor (higher factor means
smaller volume output). The new factor should be
determined using the actual weight of the delivered
fluid and the specific conditions in the user labora-
tory (temperature, atmospheric pressure etc.). This
should be done in accordance with GLP.
der derzeitige Kalibrierungsfaktor. Dieser ist bei
Auslieferung eines 20 ml Gerätes 20.00 und bei
einem 5o ml Gerät 50.00. Beim 20ml Gerät sind
die Kalibrierungsfaktoren zwischen 18.00 und
21.99, beim 50ml Gerät zwischen 45.00 und
54.99 einstellbar.
Jeweils die blikende Dezimalstelle kann durch
Drücken der rechten Taste verändert werden. Durch
Betätigen der linken Taste kann auf die nächste
Dezimalstel-le umgeschaltet werden.
Nach Auswaage vorzugsweise des Nominalvolu-
mens und Errechnen des unter Anwenderbeding-
ung tatsächlich ausgegebenen Volumens, kann die-
ses direkt als kundenspezifischer Kalibrierfaktor ein-
gegeben werden (höherer Faktor bedeutet kleinere
Ausgabevolumina).
Durch 4 sec. Drücken der linken Taste wird der ein-
gestellte Wert abgespeichert. Es erscheint kurz die
- 22 -