Set and Query the Current Range (When Using an External Current Sensor)
Syntax
Command
:CURRent:EXTRange <
External Current Sensor Range
>
Query
:CURRent:EXTRange?
Response
<External current sensor range>
<External current sensor range> = C1/C2/C5
Description
Command
Sets the external current range.
Unlike other commands such as
:CURRent:RANGe
, an error will occur if
any value other than the above <External current sensor range> is
specified.
Query
Returns the external current sensor range setting as a string.
Example
Command
:CURR:EXTR C5
Query
:CURR:EXTR?
Response
(When HEADER ON)
:CURRENT:EXTRANGE C5
(When HEADER OFF)
C5
Note
・
After you the change the range, wait a few moments until the internal circuitry
stabilizes before you read any measurement values.
・
If a range is specified, the auto range operation is turned OFF.
・
If the number of times to perform averaging is set to any value other than 1 and the
range is changed, averaging is restarted.
・
An error will occur if any value other than the above <External current sensor range> is
specified.
・
You cannot change this setting during integration, while the display is held, or when the
maximum/minimum values are being held.