62
PW3335A985-01
(9)
Instrument Display Settings
Set and Query Instrument Display Items (Normal Measurement Items)
Syntax
Command Display Area: a to d
:DISPlay[:NORMal]
<Display a>,<Display b>,
<Display c>,<Display d>
Display Area: a
:DISPlay:NORMal:A
<Display a>
Display Area: b
:DISPlay:NORMal:B
<Display b>
Display Area: c
:DISPlay:NORMal:C
<Display c>
Display Area: d
:DISPlay:NORMal:D
<Display d>
Query
Display Area: a to d
:DISPlay[:NORMal]?
Display Area: a
:DISPlay:NORMal:A?
Display Area: b
:DISPlay:NORMal:B?
Display Area: c
:DISPlay:NORMal:C?
Display Area: d
:DISPlay:NORMal:D?
Response
<Display a>,<Display b>,<Display c>,<Display d>
See
List of Measurement Item Specifications
<Display items a to d> fields.
Description
Command
Sets or queries the items to display in the instrument display areas "a" to
"d".
Example
Command
:DISP U,I,P,TIME
The instrument display area settings are as follows:
Display Area "a": Voltage (acdc)
Display Area "b": Current (acdc)
Display Area "c": Active power (acdc)
Display Area "d": Integration time
Query
:DISP?
Response
(When HEADER ON)
:DISPLAY U,I,P,TIME
(When HEADER OFF)
U,I,P,TIME
Note
・
The value displayed (instantaneous value, maximum value, or minimum value)
depends on the HOLD state.
The HOLD states and the subsequent output values are as follows:
HOLD State
Displayed Content
OFF
Instantaneous value
ON
HOLD value
Maximum value hold
Maximum value
Minimum value hold
Minimum value
・
If this command is issued while in harmonic wave display mode, use the
:DISPlay:MODE command to change to normal display mode.
・
The above-
mentioned “instantaneous value” corresponds to the averaged value while
the value is being averaged.
www.
.com
1.800.561.8187