44
PW3335A985-01
Set and Query Voltage Range Select
Syntax
Command
:VOLTage:SELect:U6V <
ON/OFF>
:VOLTage:SELect:U15V <
ON/OFF>
:VOLTage:SELect:U30V <
ON/OFF>
:VOLTage:SELect:U60V <
ON/OFF>
:VOLTage:SELect:U150V <
ON/OFF>
:VOLTage:SELect:U300V <
ON/OFF>
:VOLTage:SELect:U600V <
ON/OFF>
Query
:VOLTage:SELect:U6V?
:VOLTage:SELect:U15V?
:VOLTage:SELect:U30V?
:VOLTage:SELect:U60V?
:VOLTage:SELect:U150V?
:VOLTage:SELect:U300V?
:VOLTage:SELect:U600V?
:VOLTage:SELect:U1000V?
Response
<ON/OFF>
ON
Performs measurement using the voltage range in
question during range key operation, auto-range
operation, and auto-range integration.
OFF
Disables use of the voltage range in question during
range key operation, auto-range operation, and
auto-range integration.
Description
Command
Toggles use of the voltage range in question during auto-range operation
and auto-range integration.
Query
Returns whether the voltage range in question is used during auto-range
operation and auto-range integration.
Example
Command
:VOLT:SEL:U30V ON
Query
:VOLT:SEL:U30V?
Response
(When HEADER ON)
:VOLTAGE:SELECT:U30V ON
(When HEADER OFF)
ON
Note
・
Specifying a voltage range with the
“
:VOLTage:RANGe
”
command will cause range
select for the specified range to be turned on.
・
You cannot change this setting during integration, while the display is held, or when
the maximum/minimum values are being held.
・
Since use of the 1,000 V range is always enabled, queries will always return ON.