VT Ratio/CT Ratio
Query the VT Ratio and CT Ratio
Syntax
Query
:SCALe?
Response
<VT ratio (NR2)>,<CT ratio(NR2)>
Description
Queries the VT (PT) ratio and CT ratio setting values.
Example
Query
:SCAL?
Response
(When HEADER ON)
:SCALE:VT 2.0;CT 3.000
(When HEADER OFF)
2.0;3.000
Set and Query the VT Ratio Setting
Syntax
Command
:SCALe:VT
<VT ratio (NR2)>
Query
:SCALe:VT?
Response
<VT ratio (NR2)>
<VT ratio (NR2)> = 0.001 to 1000
Description
Command
Sets the VT (PT) ratio.
NRf numerical values are accepted.
Query
Returns the VT ratio setting in NR2 format.
Example
Command
:SCAL:VT 1.2
Query
:SCAL:VT?
Response
(When HEADER ON)
:SCALE:VT 1.200
(When HEADER OFF)
1.200
Note
・
If the number of times to perform averaging is set to any value other than 1 and the
ratio is changed, averaging is restarted.
・
Changing the setting will cause the maximum and minimum values to be reset.
・
You cannot change this setting during integration, while the display is held, or when
the maximum/minimum values are being held.
・
Instead of
:SCALe:VT(?)
, you can also use
:SCALe:PT(?)
. (Both of these commands
perform the same operation.)
Set and Query the CT Ratio Setting
Syntax
Command
:SCALe:CT
<CT ratio (NR2)>
Query
:SCALe:CT?
Response
<CT ratio (NR2)>
<CT ratio (NR2)> = 0.001 to 1000
Description
Command
Sets the CT ratio.
NRf numerical values are accepted.
Query
Returns the CT ratio setting in NR2 format.
Example
Command
:SCAL:CT 2.1
Query
:SCAL:CT?
Response
(When HEADER ON)
:SCALE:CT 2.100
(When HEADER OFF)
2.100
Note
・
If the number of times to perform averaging is set to any value other than 1 and the
range is changed, averaging is restarted.
・
Changing the setting will cause the maximum and minimum values to be reset.
・
You cannot change this setting during integration, while the display is held, or when
the maximum/minimum values are being held.