78
PW3335A985-01
Set and Query
:MEASure?
Output Items
(Instantaneous, Maximum, and Minimum Values for Voltage Data)
Syntax
Instantaneous
value
:MEASure[:NORMal]:ITEM:U:ALL <Output item>
:MEASure[:NORMal]:ITEM:U:CH1(?) <Output item>
Maximum
value
:MEASure[:NORMal]:ITEM:U_MAX:ALL <Output item>
:MEASure[:NORMal]:ITEM:U_MAX:CH1(?) <Output item>
Minimum
value
:MEASure[:NORMal]:ITEM:U_MIN:ALL <Output item>
:MEASure[:NORMal]:ITEM:U_MIN:CH1(?) <Output item>
Response
<Output item (NR1)>
128
64
32
16
8
4
2
1
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
FND
DC
AC
MN
ACDC
Description
Sets the voltage data (instantaneous value, maximum value, minimum value)
output items as numerical values between 0 and 31.
For example, specify 1 to output the AC+DC rectification or 2 to output the AC+DC
Umn rectification measurement value.
You can also output all rectification methods together at once. For example, you
can specify 3 to output both the AC+DC rectification and AC+DC Umn rectification
measurement values at the same time.
Although NRf numerical values are accepted, values to the right of the decimal are
dropped.
If no items are specified directly via a
:MEASure?
query, the measurement values
are output based on the specification of this command.
Example
Command
:MEAS:ITEM:U:CH1 1
Specifies to output the instantaneous value of the AC/DC rectified voltage.
Query
:MEAS:ITEM:U:CH1?
Response
(When HEADER ON)
:MEASURE:NORMAL:ITEM:U:CH1 1
(When HEADER OFF)
1
Set and Query
:MEASure?
Output Items
(Instantaneous, Maximum, and Minimum Values for Current Data)
Syntax
Instantaneous
value
:MEASure[:NORMal]:ITEM:I:ALL
<Output item>
:MEASure[:NORMal]:ITEM:I:CH1(?)
<Output item>
Maximum
value
:MEASure[:NORMal]:ITEM:I_MAX:ALL
<Output item>
:MEASure[:NORMal]:ITEM:I_MAX:CH1(?)
<Output item>
Minimum
value
:MEASure[:NORMal]:ITEM:I_MIN:ALL
<Output item>
:MEASure[:NORMal]:ITEM:I_MIN:CH1(?)
<Output item>
Response
<Output item (NR1)>
128
64
32
16
8
4
2
1
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
FND
DC
AC
MN
ACDC
Description
Sets the current data (instantaneous value, maximum value, minimum value)
output items as numerical values between 0 and 31.
For example, specify 1 to output the AC+DC rectification or 2 to output the
AC+DC Umn rectification measurement value.
You can also output all rectification methods together at once. For example, you
can specify 3 to output both the AC+DC rectification and AC+DC Umn
rectification measurement values at the same time.
Although NRf numerical values are accepted, values to the right of the decimal
are dropped.
If no items are specified directly via a
:MEASure?
query, the measurement
values are output based on the specification of this command.
Example
Command
:MEAS:ITEM:I:CH1 1
Specifies to output the instantaneous value of the AC/DC rectified current.
Query
:MEAS:ITEM:I:CH1?
Response
(When HEADER ON)
:MEASURE:NORMAL:ITEM:I:CH1 1
(When HEADER OFF)
1