10
Making Measurements
Attaching the Test Head and the Adapter
Attach the test head and the IM9906 Adapter to the IM9200 Test Fixture Stand.
IMPORTANT
The impedance analyzer, test head, and measurement cable are adjusted together at the time
of shipment. Use the test head and measurement cable included along with the impedance
analyzer. Confirm that the serial number* of the impedance analyzer and test head is same.
* The 9-digit serial number indicates the year (first two digits) and the month of manufacture (next
two digits). Do not remove this sticker as the number is important.
You will need:
• Test head ×1
• IM9200 Test Fixture Stand ×1
• IM9906 Adapter ×1
• 3.5 mm connector torque wrench ×1 (not included with the device)
Test head
IM9200
1
1
2
2
Metal fittings
1
Pull the metal fittings to the IM9200 Test Fixture
Stand side.
2
Place the test head on the stand and tighten the
knob.
IM9906
Notch
Measurement terminal
3
3
4
4
3
Place the notch of the IM9906 Adapter parallel to the
metal fittings, then install onto the measurement
terminal of the test head.
At this time it is stopped temporarily. Position the notch of the
IM9906 Adapter where the metal fittings can slide.
4
Fix the notch of the IM9906 Adapter by sliding the
metal fittings.
3.5 mm connector nut
5
5
5
Tighten the nut of the 3.5 mm connector by using a
torque wrench.
Recommended torque: 0.9 N·m
Tightening the nut until the handle of the wrench bends slightly
is sufficient. Do not over tighten.