17
Making Measurements
You will need:
• Flat-head screwdriver ×1 (not included with the device)
• Short plate ×1
• Tweezers ×1 (This is not included with the device. Tweezers which are thinner than the DUT size,
clean, and not magnetized are recommended.)
1
Set the reference value of correction (COMPEN) in
the impedance analyzer.
By default, the impedance analyzer is configured using the
values shown in the following table. Use the device’s short
correction and open correction defined values shown in the
following table.
OPEN
G: 0 S
,
Cp: 0 F
SHORT
Rs: 0
Ω
,
Ls: 0 H
L-side SMD
electrode
Knob
Knob
Electrodes for SMD measurement
Base plate
H-side lead
component clip
Screws
Electrodes and clips for lead component
measurement
Knob
L-side lead
component
electrode
Base plate
2
(If electrodes and clips for lead component
measurement have been installed)
Remove the H-side lead component clip and L-side
lead component electrode from the base plate.
Using a screwdriver, remove the H-side lead component
clip. Turn the knob to remove the L-side lead component
electrode.
(If the SMD measurement electrodes have been
installed)
Remove the L-side SMD electrode from the base
plate.
Turn the knob and remove the L-side SMD electrode.
Screw
Short plate
3
Using a flat-head screwdriver, install the short plate.
Tighten the short plate’s screw to the correct position. If the
screw catches partway, do not force it, but rather back it
out and then attempt to tighten it again after inserting it in a
perpendicular orientation.
Screw in correct position
Partially tightened screw
View from beside the plate
HIOKI IM9202A961-00