20
Making Measurements
6
Start the open correction in the impedance
analyzer.
1
1
22
33 44
55
6
6
When correction cannot be executed, check that
the DC measurement limit value of the impedance
analyzer is appropriate.
Measuring SMDs
(4) (3) (2) (1)
1
1
H-side
electrode
3
3
2
2
1
Choose the best L-side SMD electrode lock
position from (1) through (4) as shown in the
figure on the left based on the SMD’s dimensions
(width).
Lock position for
the L-side SMD electrode
SMD dimensions
(width)
(1)
1.6 mm to 7 mm
(2)
7 mm to 12.5 mm
(3)
12.5 mm to 18 mm
(4)
18 mm to 23 mm
2
Install the L-side SMD electrode while pressing
down on and turning the knob.
3
Pull the lever on the L-side SMD electrode toward
you to align the DUT interval with the electrode
width.