Maintenance and Service

Maintenance and Service

 CAUTION

Do not attempt to modify, disassemble, or try to repair the device. Doing so may cause 

a fire, electric shock, or injury.

Storage

If you do not intend to use the device for an extended period, store it in the carrying case.

Cleaning

 • If the device becomes dirty, wipe the device clean with a soft cloth slightly moistened with water 

or a neutral detergent.

 • The measurement accuracy and reproducibility of the device’s electrode and short plate may 

decrease due to dirt or deformation. If dirt sticks to the device, remove it by using the provided 

cleaning brush.

 • Remove the H-side lead component clip, the L-side lead component electrode, the L-side SMD 

electrode, and the short plate. Gently brush the H-side electrode and base plate.

Moving parts service life

 • The performance of some parts used in the device may degrade during the course or extended 

use. When repair is necessary, please contact your authorized Hioki distributor or reseller.

 • Service life varies with the operating environment and frequency of use. Please note that 

performance is not guaranteed for the duration of the recommended replacement cycle.

Parts

Recommended 

replacement cycle

Conditions

Base plate

10000 times

Number of times the L-side lead component 

electrode and L-side SMD electrode have slid back 

and forth

H-side lead component clip 10000 times

Number of open/close cycles

L-side lead component 

electrode

10000 times

Number of times slid back and forth over base 

plate

L-side lead component clip 10000 times

Number of times slid back and forth over base 

plate

Troubleshooting

If the device seems to be malfunctioning, contact your authorized Hioki distributor or reseller.

Transporting

Be sure to follow these precautions:

 • To avoid damaging the device, store it in the carrying case with the H-side lead component clip 

and the L-side lead component electrode connected to the base plate and with the electrode 

locked in place with the knob. Remove the short plate and L-side SMD electrode.

 • Keep the removed accessories in the original packing materials in which it was shipped, and 

store in the carrying case. Additionally, be sure to double box the device. Accidental damage 

suffered in transit is not covered by the warranty. 

 • Attach a description of the issue when sending out your device for repair.

Disposal

Dispose of the device in accordance with local regulations.

HIOKI IM9202A961-00

Содержание IM9202

Страница 1: ...dition 1 IM9202A961 00 20 04H Be sure to read this manual before using the instrument p 3 When using the instrument for the first time Troubleshooting Part Names and Functions p 6 Maintenance and Serv...

Страница 2: ...HIOKI IM9202A961 00...

Страница 3: ...t how to use the product In explaining how to use the product it assumes electrical knowledge equivalent of the knowledge possessed by a graduate of an electrical program at a technical high school Co...

Страница 4: ...Instruction manual The H side lead component clip and the L side lead component electrode are installed on the device Options sold separately The options listed below are available for the device To...

Страница 5: ...h of or serious injury to the operator CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the device or malfunction IMPORTANT...

Страница 6: ...ce Handling the Device CAUTION The DUT could pop off the device and into the air while being connected To prevent injury use gradual movements when handling clips and electrodes To avoid damage to the...

Страница 7: ...Lead wire diameter 0 5 mm to 2 mm Radial D E D Lead interval 2 mm to 26 mm E Lead length 1 2 mm or longer Lead wire diameter 0 5 mm to 2 mm SMD 2 H F G F Component length 1 6 mm to 23 mm G Component...

Страница 8: ...on The exclusive stand for fixing the test head of the impedance analyzer and the device 7 Loupe The magnifying glass provided with the IM9200 Test Fixture Stand option It is used by installing to the...

Страница 9: ...d components 10 9 2 The electrode for measurement of SMDs 13 14 13 L side SMD electrode Used when measuring SMDs 14 Lever This lever causes the L side SMD electrode to slide back and forth Short plate...

Страница 10: ...struction manual of the impedance analyzer Test head 2 Attach the test head and adapter Attach the test head and adapter to the stand to the stand Attach the test head and the IM9906 Adapter to the IM...

Страница 11: ...values fall within the allowable range See Residual parameters reference value p 24 Check items Remedy Are the electrode or the short plate worn If any of the following applies send the device for re...

Страница 12: ...9906 Adapter 1 3 5 mm connector torque wrench 1 not included with the device Test head IM9200 1 1 2 2 Metal fittings 1 Pull the metal fittings to the IM9200 Test Fixture Stand side 2 Place the test he...

Страница 13: ...f the internal conductor rotates it may cause damage to the electrode When the calibration cannot be executed Confirm that the calibration kit and the IM9906 Adapter are connected properly Check that...

Страница 14: ...each other and the calibration kit and adapter may get damaged 6 Execute open calibration by using the impedance analyzer 2 2 1 1 0 0 3 3 4 4 Nut Short calibration kit 9 9 7 Remove the open calibratio...

Страница 15: ...in figure connect by turning only the nut of load calibration kit while pressing the top part of the load calibration kit If the top part of the load calibration kit is rotated the internal conductors...

Страница 16: ...1 Lower the screw thread by turning the nut of the IM9906 Adapter 2 2 2 2 3 3 2 Loosen the stand fixing knob 3 Place the device on the IM9200 Test Fixture Stand As shown in figure slide the notch part...

Страница 17: ...he error caused by the phase shift occurring between the calibration reference surface and the measurement DUT s connection surface is compensated Set the electrical length correction to 0 mm in the i...

Страница 18: ...he difference in impedance between the DUT and the short device The short device s impedance will affect measured values Using a fixed reference without regard to the size of the DUT The value after s...

Страница 19: ...easurement Knob L side lead component electrode Base plate 2 If electrodes and clips for lead component measurement have been installed Remove the H side lead component clip and L side lead component...

Страница 20: ...r When the correction cannot be executed Confirm that the IM9906 Adapter is connected properly Check that the DC measurement limit value of the impedance analyzer is appropriate 1 1 2 2 3 3 4 4 5 5 0...

Страница 21: ...screwdriver install the H side lead component clip 2 Gently tighten the knob that holds the L side lead component electrode in place Edge 3 3 H side electrode Marker 3 Slide the L side lead component...

Страница 22: ...hoose the best L side SMD electrode lock position from 1 through 4 as shown in the figure on the left based on the SMD s dimensions width Lock position for the L side SMD electrode SMD dimensions widt...

Страница 23: ...the plate Catch the other protruding part on the leading edge of the L side SMD electrode View from the side with the jig installed SMD open correction jig If you re unable to secure electrode positi...

Страница 24: ...er on the L side electrode to sandwich the component between the H side electrode and the L side SMD electrode Verify that the edge of the L side SMD electrode is in contact with the base plate Proper...

Страница 25: ...ation Connecting part to the DUT 2 terminal structure Lead components The lead component s leads are pinched between the device s electrodes and clips SMDs The SMD is pinched between the device s elec...

Страница 26: ...y connected to the H side electrode The H side clip is present 1 2 nH 1 1 nH 30 MHz 1 2 nH 0 8 nH 100 MHz 1 2 nH 0 7 nH 300 MHz 600 MHz Residual parameters Short inductance L side SMD electrode Condit...

Страница 27: ...d for the duration of the recommended replacement cycle Parts Recommended replacement cycle Conditions Base plate 10000 times Number of times the L side lead component electrode and L side SMD electro...

Страница 28: ...26 Maintenance and Service HIOKI IM9202A961 00...

Страница 29: ...HIOKI IM9202A961 00...

Страница 30: ...HIOKI IM9202A961 00...

Страница 31: ...HIOKI IM9202A961 00...

Страница 32: ...HIOKI IM9202A961 00...

Отзывы: