24
Specifications
Residual parameters (reference value)
Residual parameters
Short inductance
(Short plate)
Conditions
•Inductance value as seen from the 7 mm connector calibration surface (Electrical
Length Correction 0 mm)
Measurement parameter Ls
•Short plate connection
1.6 nH
±
1.1 nH: 30 MHz
1.6 nH
±
0.8 nH: 100 MHz
1.6 nH
±
0.7 nH: 300 MHz, 600 MHz
Residual parameters
Short inductance
(L-side lead component
electrode)
Conditions
•Inductance value as seen from the 7 mm connector calibration surface (Electrical
Length Correction 0 mm)
Measurement parameter Ls
•The L-side lead component electrode is directly connected to the H-side electrode.
•The H-side clip is present.
1.2 nH
±
1.1 nH: 30 MHz
1.2 nH
±
0.8 nH: 100 MHz
1.2 nH
±
0.7 nH: 300 MHz, 600 MHz
Residual parameters
Short inductance
(L-side SMD electrode)
Conditions
•Inductance value as seen from the 7 mm connector calibration surface (Electrical
Length Correction 0 mm)
Measurement parameter Ls
•The L-side SMD electrode is directly connected to the H-side electrode.
1.2 nH
±
1.1 nH: 30 MHz
1.2 nH
±
0.8 nH: 100 MHz
1.2 nH
±
0.7 nH: 300 MHz, 600 MHz
Residual parameters
Open capacitance
Conditions
•Capacitance value as seen from the 7 mm connector calibration surface (Electrical
Length Correction 0 mm)
Measurement parameter Cp
•Neither the L-side electrode nor the H-side clip is present.
1.3 pF
±
0.3 pF (30 MHz)
1.3 pF
±
0.2 pF (100 MHz, 300 MHz, 600 MHz)