21
Making Measurements
SMD open
correction jig
Protruding part
Leading edge of
the L-side SMD electrode
Gap between the top
surface of the electrode
and the plate
4
4
Protruding part
4
Install the SMD open correction jig and lock the
electrode in place.
Firmly push the protruding part of the SMD open correction
jig into the gap between the top surface of the L-side SMD
electrode and the plate. Catch the other protruding part on
the leading edge of the L-side SMD electrode.
View from the side with the jig installed
SMD open correction jig
If you’re unable to secure electrode position
sufficiently, you’ll need to replace the SMD open
correction jig. Please contact your authorized Hioki
distributor or reseller
.
5
Start the open correction in the impedance
analyzer.
1
1
22
33 44
55
6
6
When correction cannot be executed, check that
the DC measurement limit value of the impedance
analyzer is appropriate.
6
Remove the SMD open correction jig.