50
Single-phase Active Energy Measurement (Integrated Measurement) [Setting Wh]
4
Set the instrument to
power integration mode.
5
Start integration.
•
Range is fixed when integration starts.
Change the range before starting to
integrate a large power quantity or for an
extended period of time.
•Only the positive (consumption) active
power is added. Negative (generation)
active power is not added.
6
Stop integration.
Single-phase active power
Single-phase active energy
Elapsed time
•When the
HOLD
key is pressed during integration stop, the integrated energy clears and
returns to the display shown in Step 4.
•The measured values are automatically stored just before the instrument turns off due to low
battery voltage. Next time the instrument is turned on, the saved values will be displayed. (The
measured values can be cleared by pressing the
HOLD
key.)
HIOKI CM3286C961-00