78
Specifications
Input
,
output
,
and measurement specifications
Basic specifications
Measurement items
AC current RMS value/AC current peak value (no polarity)/AC current frequency
AC voltage RMS value/AC voltage peak value (no polarity)/AC voltage frequency
Single-phase active power/Single-phase apparent power/Single-phase reactive
power/Single-phase power factor/Single-phase zero-cross phase angle
Balanced 3-phase active power/Balanced 3-phase reactive power/Balanced
3-phase apparent power/Balanced 3-phase power factor/Balanced 3-phase
zero-cross phase angle
Single-phase active energy (only positive values added)/Phase detection
Maximum input current
As per the frequency derating characteristics (p. 79).
(up to 200 Hz
,
600 A or less; above 200 Hz
,
120000 A
Hz or less)
Maximum rated voltage
to terminal
600 V AC
Maximum rated voltage
to earth
600 V AC (Measurement category IV)
,
1000 V AC (Measurement category III)
Anticipated transient overvoltage 8000 V
Maximum measuring
voltage
600 V AC
Measurement method
True RMS measurement with digital sampling
Measurement terminals
COM terminal
,
V terminal
Input impedance
1 M
Ω
or greater