54
Single-phase Energy Meter Comparison Function [Setting Wh]
4
Set the instrument
to power
integration mode.
Single-phase active power (measured value
)
5
Start integration.
Energy (theoretical value)
Elapsed time
•Only the positive (consumption) active power is added. Negative
(generation) active power is not added.
•
Range is fixed when integration starts. Change the range before starting
to integrate a large power quantity or for an extended period of time.
6
Stop integration.
Energy ratio (Measured value / theoretical value)
Energy difference
(Measured value − theoretical value)
Energy (Measured value)
When the
HOLD
key is pressed during integration stop, the integrated energy clears and returns
to the display shown in Step 4.