94
Equations
Equations
(1) Single‑phase power measurement
Apparent power
S
U
RMS
∙
I
RMS
•
The active power
P
has no sign
during consumption and a negative
sign during generation.
•
Due to the effects of measurement
error
,
S
=|
P
| and
Q
=0 are used when
S
<|
P
|.
Reactive power
Q
2
2
P
S
−
Power factor
PF
S
P
(2) Balanced 3‑phase/3‑wire power measurement
Balanced 3‑phase/3‑wire
zero‑cross phase angle
φ
(3P3W)
φ
−30°
•
The symbol
φ
represents the
zero-cross phase angle of the
voltage
U
12
and the current
I
1
.
•
The symbol
S
represents
the apparent power of the
line voltage
U
12
and the wire
current
I
1
.
Balanced 3‑phase/3‑wire
power factor
PF
(3P3W)
{
}
)
3
3
(
cos
W
P
φ
Balanced 3‑phase/3‑wire
active power
P
(3P3W)
[W]
S
W
P
⋅
⋅
)
3
3
(
3
PF
Balanced 3‑phase/3‑wire
apparent power
S
(3P3W)
[VA]
S
⋅
3
Balanced 3‑phase/3‑wire
reactive power
Q
(3P3W)
[var]
2
)
3
3
(
2
)
3
3
(
W
P
W
P
P
S
−
HIOKI CM3286C961-00
Содержание CM3286-50
Страница 2: ...HIOKI CM3286C961 00...
Страница 6: ...iv Contents HIOKI CM3286C961 00...
Страница 114: ...108 Troubleshooting HIOKI CM3286C961 00...
Страница 115: ...HIOKI CM3286C961 00...
Страница 116: ...HIOKI CM3286C961 00...
Страница 117: ...HIOKI CM3286C961 00...
Страница 118: ...www hioki com HIOKI CM3286C961 00...