107
Troubleshooting
Display
Buzzer
Cause
Solution
e.g.: for current
measurement
Flashes
red
Intermittent
sound
A current or voltage
exceeding the
maximum input
was input to the
instrument.
Stop measurement
immediately as the
current or voltage cannot
be measured by the
instrument. For current
measurement
,
the
optional 9290-10 can be
used to measure currents
of up to 1000 A AC.
When manual range is
6 A and 60 A range
,
this
warning display will not
appear.
e.g.: for current
measurement
Lights
red
–
A current or voltage
exceeding the range
was input while using
a manual range.
Change the measurement
range or select the AUTO
range.