
5.2 Signal Descriptions
77
5
Chap
ter
5
Ex
ter
nal
Cont
rol
(EXT
I
/O)
LOAD0 to
LOAD6
Select a Panel No. to load and apply a TRIG signal to load the selected Panel
No. and measure. LOAD0 is the LSB, and LOAD6 is the MSB.
When a TRIG signal is applied, if LOAD0 through LOAD6 are unchanged from
the previous trigger event, panel settings are not loaded. In this case, using
external triggering, one measurement is taken as usual when the TRIG signal is
applied.
• At least 70 ms is required for the settings to change after executing a Panel
Load (the actual time depends on the particular function, range and sampling
rate).
• When set to external trigger mode, one measurement is taken upon load com-
pletion.
TRIG
When the external trigger, one measurement is taken each time the TRIG signal
transitions from High to Low.
This trigger signal is ignored when internal triggering is enabled.
Trigger functions are also available for statistical calculation, recording to mem-
ory and output of measured values (valid also with internal triggering).
CAL
When manual self-calibration is selected with FAST or MEDIUM sampling rate,
self-calibration begins when the CAL signal transitions from High to Low.
Self-calibration takes about 176 ms (power supply frequency: 50 Hz) or about
151 ms (ower supply frequency: 60 Hz).
When SLOW sampling is selected, the CAL signal is ignored.
See
Input Signals
Panel
No.
LOAD6
LOAD5
LOAD4
LOAD3
LOAD2
LOAD1
LOAD0
0
0
0
0
0
0
0
1
0
0
0
0
0
0
1
2
0
0
0
0
0
1
0
3
0
0
0
0
0
1
1
4
0
0
0
0
1
0
0
5
0
0
0
0
1
0
1
6
0
0
0
0
1
1
0
7
0
0
0
0
1
1
1
8
0
0
0
1
0
0
0
...
122
1
1
1
1
0
1
0
123
1
1
1
1
0
1
1
124
1
1
1
1
1
0
0
125
1
1
1
1
1
0
1
126
1
1
1
1
1
1
0
1
1
1
1
1
1
1
0: HIGH: Open or from 5 V to 24 V
1: LOW: 0 V to 0.9 V
When a TRIG signal is applied with LOAD0 to LOAD6 set to all 1's or all 0's,
no Panel Load occurs.
HIOKI BT3564A961-02
Содержание BT3564
Страница 2: ...HIOKI BT3564A961 02...
Страница 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Страница 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Страница 89: ...5 4 Internal Circuitry 83 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3564A961 02...
Страница 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Страница 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Страница 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Страница 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Страница 206: ...Index 4 HIOKI BT3564A961 02...
Страница 207: ...HIOKI BT3564A961 02...
Страница 208: ...HIOKI BT3564A961 02...
Страница 209: ...HIOKI BT3564A961 02...
Страница 210: ...HIOKI BT3564A961 02...