9.4 Functions
170
9.4 Functions
Zero-adjustment function
Enable or cancel zero-adjustment
Zero-adjustment setting: ON/OFF
Cancel zero-adjustment: Turn off the zero-adjustment to clear all zero-adjust-
ment data.
Zero-adjustment range
Resistance measurement: -1000 to 1000 counts
Voltage measurement: -1000 to 1000 counts
Self calibration
Calibration mode: AUTO/MANUAL
AUTO: executes automatically once every 30 minutes
MANUAL: executes manually by EXT I/O signal or remote command
* When sampling is set to SLOW, the instrument performs self-calibration
before each measurement.
Trigger function
Trigger source: Internal/External
Delay function
Delay setting: ON/OFF
Delay time: 0 to 9.999 sec
Averaging function
Averaging setting: ON/OFF
No. of samples to average: 2 to 16 times
Comparator function
Comparator function setting: ON/OFF
Comparator setting
Comparator mode: HIGH, LOW/REF, %
Upper and lower limit value: 0 to 99999 (Resistance)/0 to 999999 (Voltage)
Reference value: 0 to 99999 (Resistance)/0 to 999999 (Voltage)
%: 0.000% to 99.999% (percentage range setting applies to both positive
and negative values)
Comparator judgment beeper mode: OFF/HIGH, LOW/IN/ALL
Operating mode: AUTO/MANUAL
* Measurement value or statistical 3
(population standard deviation
×
3)
can be set as upper/lower limit or reference values.
Decision
Judgment result: Hi/IN/Lo (resistance and voltage judged independently)
PASS/FAIL judgement Calculates the logical AND of resistance and voltage
judgment results and outputs a PASS/FAIL judgment (EXT I/O output).
Measurement fault value judgments:
OF; Hi judgment
-OF; Lo judgment
Measurement fault; Not judged (no judgment result)
Statistical calculation
function
Statistical calculation setting: ON/OFF/Clear
Auto-clear after printing statistical data
Calculations: Total data counts, valid data counts, maximum, minimum,
average, standard deviation, population standard deviation and process
capability indices (Cp and CpK)
Calculations trigger: Statistical calculation of measured values initiated by
EXT I/O signals, key or remote command
HIOKI BT3564A961-02
Содержание BT3564
Страница 2: ...HIOKI BT3564A961 02...
Страница 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Страница 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Страница 89: ...5 4 Internal Circuitry 83 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3564A961 02...
Страница 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Страница 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Страница 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Страница 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Страница 206: ...Index 4 HIOKI BT3564A961 02...
Страница 207: ...HIOKI BT3564A961 02...
Страница 208: ...HIOKI BT3564A961 02...
Страница 209: ...HIOKI BT3564A961 02...
Страница 210: ...HIOKI BT3564A961 02...