4.2 Trigger Function
58
Specify the delay from the moment a trigger is applied to the start of measurement. By using this function,
even when a trigger is applied immediately after connecting a test object, the start of measurement can be
delayed to allow sufficient time for the measurement value to stabilize. Trigger delay can be set with 1 ms res-
olution from 0.000 to 9.999 seconds.
Trigger Delay Settings
1
(The SHIFT indicator lights up.)
The Trigger Delay setting display appears.
2
Select
ON
.
3
The numerals indicating the trigger delay blink.
4
Set the trigger delay.
5
Applies the setting and returns to the Measurement display.
To cancel the settings:
SHIFT
key
(Main display)
(Sub display)
The current setting blinks.
(Sub display)
(Sub display)
Or numeric keypads
Disabling the Trigger Delay Function
1
(The SHIFT indicator lights up.)
The Trigger Delay setting display appears.
2
Select
OFF
.
3
The Trigger Delay is disabled.
(Sub display)
HIOKI BT3564A961-02
Содержание BT3564
Страница 2: ...HIOKI BT3564A961 02...
Страница 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Страница 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Страница 89: ...5 4 Internal Circuitry 83 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3564A961 02...
Страница 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Страница 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Страница 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Страница 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Страница 206: ...Index 4 HIOKI BT3564A961 02...
Страница 207: ...HIOKI BT3564A961 02...
Страница 208: ...HIOKI BT3564A961 02...
Страница 209: ...HIOKI BT3564A961 02...
Страница 210: ...HIOKI BT3564A961 02...