8.6 Message Reference
147
8
Chapt
er 8
RS-232
C
/GP-I
B
In
te
rf
aces
Continuous Measurement Setting
Syntax
Command
:INITiate:CONTinuous
<1, 0, ON or OFF>
Query
:INITiate:CONTinuous?
Response
<ON or OFF>
ON ........ Continuous Measurement Enabled
OFF....... Continuous Measurement Disabled
Description
Command Sets continuous measurement.
Query
Queries the continuous measurement setting.
Example
Command
:INIT:CONT OFF
Disables continuous measurement.
Query
Response
:INIT:CONT?
ON
Enables continuous measurement.
Note
• Continuous Measurement Enabled:
After measurement, enters the Trigger Wait State. When the trigger source set-
ting is IMMediate, the next trigger occurs immediately (the Free-Run State).
• Continuous Measurement Disabled:
After measurement, enters the Idle State instead of the Trigger Wait State.
• Triggering is ignored in the Idle State. Executing :INITiate[:IMMediate] enables
the Trigger Wait State.
• Continuous measurement is enabled upon exit from the Remote State.
Trigger Wait Setting
Syntax
Command
:INITiate[:IMMediate]
Description
Command Switches triggering from the Idle State to the Trigger Wait State.
Example
Command Disable continuous measurement, and read one value for each trigger
event
Send
:TRIG:SOUR IMM
.... Trigger immediately when entering Trigger
Wait State
:INIT:CONT OFF
.... Disables continuous measurement
:INIT
........................... Enable Trigger Wait Trigger immediately
upon :TRIG:SOUR IMM
:FETC?
........................ Fetch measured value
Response
2.1641E+0
................. Measured value is 2.1641
Error
• An execution error occurs when continuous measurement is enabled (
:INI-
TIATE:CONTINUOUS ON
).
Note
• When the trigger source is IMMediate, triggering occurs immediately before
entering the Idle State.
• When the trigger source is EXTernal, the Trigger Wait State is enabled to wait
for an external trigger, and when a trigger occurs, one measurement is taken
before entering the Idle State.
HIOKI BT3564A961-02
Содержание BT3564
Страница 2: ...HIOKI BT3564A961 02...
Страница 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Страница 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Страница 89: ...5 4 Internal Circuitry 83 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3564A961 02...
Страница 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Страница 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Страница 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Страница 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Страница 206: ...Index 4 HIOKI BT3564A961 02...
Страница 207: ...HIOKI BT3564A961 02...
Страница 208: ...HIOKI BT3564A961 02...
Страница 209: ...HIOKI BT3564A961 02...
Страница 210: ...HIOKI BT3564A961 02...