8.6 Message Reference
138
Query the Memory Data Count
Syntax
Query
:MEMory:COUNt?
Response
<Memory data count>
<Memory data count>
= 0 to 400
(NR1)
Example
Query
:MEM:COUN?
Response
5
Query (Download) Memory Data
Syntax
Query
:MEMory:DATA? [STEP]
Response
<Memory data No. (NR1)>,<Measured resistance (NR3)>,<Measured
voltage (NR3)>
Memory data values are returned as data objects.
If [STEP] is omitted, all memory data objects are returned continu-
ously.
Example
Query
:MEM:DATA?
Example
Response
1, 290.60E-3, 1.3924E+0
2, 290.54E-3, 1.3924E+0
3, 290.50E-3, 1.3923E+0
4, 290.43E-3, 1.3923E+0
5, 290.34E-3, 1.3924E+0
END
Query
:MEM:DATA? STEP
Response
1, 290.60E-3, 1.3924E+0
N
(Sent from PC)
2, 290.54E-3, 1.3924E+0
N
(Sent from PC)
3, 290.50E-3, 1.3923E+0
N
(Sent from PC)
4, 290.43E-3, 1.3923E+0
N
(Sent from PC)
5, 290.34E-3, 1.3924E+0
N
(Sent from PC)
END
Note
• Stored memory data objects are returned continuously, or one data object at a
time. The “END” character is returned as the last data object. When the “STEP”
parameter is specified, one data object is returned at a time. Sending “N” to the
instrument after receiving the data causes the next data object to be returned.
The memory index is an unsigned three-digit integer. Refer to “Measurement
Value Formats” for format details of returned measurement values.
• A terminator is appended to the end of each returned memory data object.
When sending “N” from the PC or other device, a terminator is required.
See
• Measured values are stored in memory when pressing the
TRIG
key, applying a
signal to the TRIG EXT I/O connector or sending the
TRG
command (while
the Memory function is enabled). Up to 400 data objects can be stored. When
the memory is full, additional measurement data is not stored.
• When the Memory function is enabled, auto-ranging is disabled.
• When the measurement mode is set to
or V, a measurement error value will
be returned for functions that are not being measured.
HIOKI BT3564A961-02
Содержание BT3564
Страница 2: ...HIOKI BT3564A961 02...
Страница 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Страница 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Страница 89: ...5 4 Internal Circuitry 83 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3564A961 02...
Страница 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Страница 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Страница 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Страница 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Страница 206: ...Index 4 HIOKI BT3564A961 02...
Страница 207: ...HIOKI BT3564A961 02...
Страница 208: ...HIOKI BT3564A961 02...
Страница 209: ...HIOKI BT3564A961 02...
Страница 210: ...HIOKI BT3564A961 02...