4.1 Comparator Function
49
4
Chapt
er
4
Appl
ie
d Measur
ement
15
Connect a test object and judge the measured value.
In the
V mode, you can verify comparator settings by pressing the
VIEW
key.
See
" Switching Between Measurement Value and Comparator Setting Displays" (
The measured voltage value is displayed as its relative percentage offset
from the reference value (%)
The measured resistance value is displayed as
its relative percentage offset from the reference
value (%)
Judgment result
Measured resistance value - Reference value
=
Reference value
x 100
Relative
percentage
Upper threshold value of setting range < Measured value
Lower threshold value of setting range
Measured value
Upper
threshold value of setting range
Measured value < Lower threshold value of setting range
The instrument can also base judgments on the absolute value of measured
voltage value (to prevent Lo judgments when the positive and negative terminals
are connected backwards).
See
Configuring the Absolute Value Judgment Function (Voltage)
HIOKI BT3564A961-02
Содержание BT3564
Страница 2: ...HIOKI BT3564A961 02...
Страница 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Страница 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Страница 89: ...5 4 Internal Circuitry 83 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3564A961 02...
Страница 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Страница 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Страница 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Страница 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Страница 206: ...Index 4 HIOKI BT3564A961 02...
Страница 207: ...HIOKI BT3564A961 02...
Страница 208: ...HIOKI BT3564A961 02...
Страница 209: ...HIOKI BT3564A961 02...
Страница 210: ...HIOKI BT3564A961 02...