ii
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
C
AREGIVER
C
ONTROL
P
OD
NOTE:
There are two scale systems available for the bed, refer to the images and page numbers shown below
to determine which instructions apply to your scale system
Item
Feature
Page
A
Weigh
—press this control to take a weight reading.
B
Zero
—with the Enable key active, press this control to zero the bed.
C
Raise the Bed
—the indicator on this control flashes when you attempt to use a
scale control while the bed is in transport position (power drive wheels on the
ground). Press and hold the control until the bed is no longer in the transport
position (the indicator will turn off, and you will hear a single beep).
D
Enable key
—when active lets you zero the bed, set the Bed Exit alert, set the 30°
Head Angle alert, and adjust the alert volume and tone.
E
Bed Exit: Patient Position mode
—with the Enable key active, press this control
to set the Bed Exit alert to notify you when the patient begins to move.
F
Bed Exit: Bed Exiting mode
—with the Enable key active, press this control to set
the Bed Exit alert to notify you when the patient attempts to get out of the bed.
G
Bed Exit: Out-of-Bed mode
—with the Enable key active, press this control to set
the Bed Exit alert to notify you when the patient has left the bed.
H
Alert Silence
—when a Bed Exit alert is armed, press this control to stop
monitoring the patient movement temporarily (30 seconds).
I
Alert Volume—with the Enable key active, a patient on the bed, and a Bed Exit
mode armed, press and release this control to adjust the volume of the alert.
J
30° Head Angle Alert
—with the Enable key active, press this control to be
notified when the bed’s head section goes below 30°.
K
Digital display
—shows the head section angle or patient weight.
L
Magnification (Mag) mode
—shows the weight in the nearest 0.5 kg.
M
Frame Setup
—puts the bed in the correct position to zero the scale or weigh the
patient.
N
Release Brake
—this indicator flashes if the brakes are set and you attempt to zero
the scale.The brakes need to be released to accurately zero the scale.
O
Scale Reference
—this indicator comes on solid when the bed is correctly in the
Scale Reference Position.
P
Zero instructions—shows the sequential steps to zero the scale.
Scale A
Scale B
Refer to page 43.
Refer to page 45.