Preventive Maintenance
63
4
Verify APL Valve accuracy.
a
Attach a breathing circuit the patient module.
b
Connect the y-piece to the hose end where the reservoir bag normally is connected.
c
Set a fresh gas flow of 5 l/min.
d
Set the APL valve to each pressure graduation (10, 20, 30, 40 and 50 Pa x 100 (mbar;
cmH
2
O))
e
Compare the valves on the pressure gauge with the valve settings. Deviations from pressure
indication and setting must not exceed ±20% of the set value or 10 Pa x 100 (10 mbar; 10
cmH
2
O), whichever is greater.
4.3.4.2 Alarm Tests
1
Turn on the and ensure adequate driving gas is present.
2
Perform the Compliance test.
3
Verify the
V
TE
min (low alarm), Peak pressure (high and low alarms), M.Vol (low alarm) and O
2
(high and low alarms).
a
Ventilate a test lung in the CMV mode.
b
Set each parameter’s alarm limit one at a time, to violate it’s high and low alarm setting.
c
Verify that alarm indications function.
Fig. 20 Alarm screen
Содержание Modular+
Страница 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Страница 12: ...Theory of Operation 12 This page intentionally left blank...
Страница 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Страница 90: ...Preventive Maintenance 90...
Страница 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Страница 102: ...Preventive Maintenance 102 This page intentionally left blank...
Страница 112: ...Order Information 112 034 2950 Silicone tube 6 x 2 for suction 323 0362 Tube cable holder for holding device...
Страница 113: ...Order Information 113 This page intentionally left blank...