![Heyer Modular+ Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/heyer/modular/modular_service-manual_2128284050.webp)
Maintenance and Calibration
50
3.5 Required Tools
Item
Number
Part
Number
Description
Specification
1
-
Operation manual
HEYER specification
2
-
Service manual
HEYER specification
3
900-4726 Calibration pump, Pressure sensors
0 – 80 cmH
2
O
4
-
Safety analyzer
For electrical tests according to IEC-
601-1,B
5
-
Digital Volt Meter
6
-
Calculator
Standard four function
7
900-4715 Temperature simulator
HEYER specification
8
-
Mass flow meter
0 – 80 l/min ± 2% of reading
9
-
Surface cleaner
Standard hospital grade
10
340-3000 Gasket set
HEYER specification
11
460-0610 Driving gas test adapter with 1 meter
tubing
HEYER specification
12
603-4000 Test lung
Adult size
13
-
Test lung
Neonate size
14
-
High pressure gauge
0 – 2.5 kPa x 100(0 – 2.5 bar;
0 – 38 PSI)
15
-
Low pressure gauge
0 – 250 Pa x 100(0 – 250 mbar;
0 – 3.8 PSI)
17
-
Hand tools, calibration tool, flat screw
driver, socket and Allen wrench set
Metric standard
3.6 Disassembly instructions
3.6.1 Connecting and disconnecting vaporizers
1. To connect the vaporizer, place it carefully onto the vaporizer manifold. The manifold is Selectatec
type.
2. Lock the vaporizer with the locking lever ¼ turn to the fully locked position.
3. Only vaporizers with interlock-system may be used with this system.
4. After each exchange of a vaporizer, perform a leak test.
3.6.2 Removing the compressed gas tanks (PIN-Index)
1. Ensure that the tank is closed
2. Rotate the hand tight knob that retains the tank against the port seat.
3. Flip the swiveling yoke collar 180 degrees.
4. Grip the tank firmly and remove it from the yoke.
5. Connect a new tank in reverse order, ensuring a single tank washer is present on the yoke port
seat.
Содержание Modular+
Страница 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Страница 12: ...Theory of Operation 12 This page intentionally left blank...
Страница 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Страница 90: ...Preventive Maintenance 90...
Страница 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Страница 102: ...Preventive Maintenance 102 This page intentionally left blank...
Страница 112: ...Order Information 112 034 2950 Silicone tube 6 x 2 for suction 323 0362 Tube cable holder for holding device...
Страница 113: ...Order Information 113 This page intentionally left blank...