Maintenance and Calibration
49
3.4.3 Symptoms of Fuse Failures
The following table shows the reaction to a failed fuse on the power supply board. The fuses F2, F3,
F4, F5 and F6 have been removed from the power supply module and are therefore subject of the
production date.
FUSE
Failure during working
Failure during start
F1
Display is black; module 1 and 2 do not
work
Display is black; module 1 and 2 do
not boot
F7
alarm „Solenoid Valves faulty“, „Fan
faulty“ and „Ambient Air Intake“; real
time curve shows a zero line
after booting the message „Selected
mode is not Standby“ is shown;
module 2 does not boot; fan does not
work
F8/F9
alarm „AC power lost, using battery“
after the machine has booted alarm
„AC power lost, using battery“ is
shown
F10/F11
During the ventilation there is no failure;
after you have turned the selector
switch to Standby the message „Insert
patient module“ is shown; during using
battery the message disappeared
after the machine has booted the
message„Insert patient module“ is
shown; during using battery the
message disappeared
Содержание Modular+
Страница 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Страница 12: ...Theory of Operation 12 This page intentionally left blank...
Страница 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Страница 90: ...Preventive Maintenance 90...
Страница 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Страница 102: ...Preventive Maintenance 102 This page intentionally left blank...
Страница 112: ...Order Information 112 034 2950 Silicone tube 6 x 2 for suction 323 0362 Tube cable holder for holding device...
Страница 113: ...Order Information 113 This page intentionally left blank...