Repair Information
40
2.15.2.7.7 Compliance Test Patient Module
The compliance test patient module is divided in 3 phases.
Phase 1. Flow into the re-breathing system and test of flow sensors.
Valve
MV 1
MV2
MV3
MV4
Activity
ON
OFF
ON
ON
Pressure on port
5, Yes
1, No
6, Yes
no sup. of MV2
Phase 2. Pressure increase up to test pressure 40 cmH
2
O, calculation of system compliance.
Valve
MV 1
MV2
MV3
MV4
Activity
ON
ON
ON
ON
Pressure on port
5, Yes
1, Yes
6, Yes
supply of MV2
Phase 3. Test of pressure loss.
Valve
MV 1
MV2
MV3
MV4
Activity
ON
ON
ON
ON
Pressure on port
5, Yes
1, Yes
6, Yes
supply of MV2
2.15.2.7.8 Leak Test Patient Module and Fresh Gas Module
Valve
MV 1
MV2
MV3
MV4
Activity
ON
OFF
OFF
OFF
Pressure on port
5, Yes
1, No
6, No
no sup. of MV2
Содержание Modular+
Страница 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Страница 12: ...Theory of Operation 12 This page intentionally left blank...
Страница 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Страница 90: ...Preventive Maintenance 90...
Страница 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Страница 102: ...Preventive Maintenance 102 This page intentionally left blank...
Страница 112: ...Order Information 112 034 2950 Silicone tube 6 x 2 for suction 323 0362 Tube cable holder for holding device...
Страница 113: ...Order Information 113 This page intentionally left blank...