Preventive Maintenance
56
CAUTION:
The spring in the top of the APL valve may not be stressed. After removal,
ensure that the APL valve is in position “CMV” and place to one side, taking
care that the spring is not unduly loaded.
CAUTION:
Use cleaning agent sparingly. Excess fluid could enter the machine, causing
damage.
4.3 Test Procedure
4.3.1 General
1
Ensure that the Service Manual and Operation Manual are present.
2
Ensure there is no external damage to machine and accessories by visual inspection.
3
Ensure all accessories are complete and functional by visual inspection.
4.3.2 Inspecting/replacing consumable parts
1
Remove the patient module from the
a
Turn on the and ensure adequate driving gas is present.
b
Remove the CO
2
absorber dome.
c
Remove the bag in bottle system.
d
Remove the O
2
cell cable from the O
2
cell.
e
Remove the scavenger hose.
f
Remove the breathing circuit.
g
Turn the locking handle 180° counterclockwise
h
Pull out the patient module.
i
Verify that the patient module window changes from Unlock to Unlocked.
Содержание Modular+
Страница 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Страница 12: ...Theory of Operation 12 This page intentionally left blank...
Страница 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Страница 90: ...Preventive Maintenance 90...
Страница 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Страница 102: ...Preventive Maintenance 102 This page intentionally left blank...
Страница 112: ...Order Information 112 034 2950 Silicone tube 6 x 2 for suction 323 0362 Tube cable holder for holding device...
Страница 113: ...Order Information 113 This page intentionally left blank...