Evolution™ Electric Bed LI156Ex
14
QD1532A(2)
USER MANUAL
IN
S
T
RUCT
IO
NS
F
O
R
U
S
E
4.3
ELECTRICAL FUNCTION LOCK-OUT
See Safety – General precautions for use, section 7.1
The electrical function lock-out panel can
be used to lock out or to release the
three power-driven functions either singly
or simultaneously.
Once a function is locked out, it can no
longer be operated using the siderail
controls, variable height pedal or hand
pendant control.
Controls released
Controls locked out
To lock out an electrical function:
- Switch the relevant control switch to lock:
.
To release an electrical function:
- Switch the relevant control switch to unlock:
.
IMPORTANT: Locking out the Knee Section adjustment control will also
lock out the Auto-contour feature (see sect. 4.5.3.).
Locking out the height adjustment control will also lock out the Reverse
Trendelenburg option when the bed is in a raised position.
Locking out the Head Section does not lock out the Auto-contour Knee
Section.
When the patient is restrained by the straps, the electrical functions must
be locked out.