- Never try to move the bed by pulling on the power cord
- It is advisable to remove the patient helper when transporting the bed
and bedside cabinet through low doorways (< 2.4 m).
Two people should move the bed (one at each end) in the following
situations:
- Moving the bed on a slope
- Foot end steer wheel
- Transporting with bedside cabinet
- Moving the bed with a heavy load (heavy patient, accessories
fitted, water mattress, etc.).
To move the bed:
- Grip the footboard or frame with
both hands
- Use the brake and steer pedal
to unlock the wheels
- Push the bed, steering with the
footboard or frame.
When moving in a straight line, position the pedal on "STEER" and push the
bed using the endboard opposite the steering wheel (see section 4.7.1).
4.8
EQUIPOTENTIAL TERMINAL
When direct intravascular or intracardiac connections are in use, the
electrical potentials of all the unprotected metal bed parts need to be
equalised.
To equalise potentials, connect the equipotential cable (ref. AC968A) to the
connection terminal.
PLEASE NOTE: Failure to connect the equipotential cable may result in
injury to the patient through electric shock.