Evolution™ Electric Bed LI156Ex
QD1532A(2)
17
USER MANUAL
IN
S
T
RUCT
IO
NS
F
O
R
U
S
E
Trendelenburg/Reverse Trendelenburg with or without power supply
(mains or battery)
Bed position:
high
Bed position: low
Bed position:
mid-height
Power on
Trendelenburg
/ Reverse
Trendelenburg
possible
Raise bed to use
Trendelenburg /
Reverse
Trendelenburg
tilting
Trendelenburg
/ Reverse
Trendelenburg
possible
Power off
(mains and
battery)
Trendelenburg
possible
Reverse
Trendelenburg
possible
Trendelenburg
/ Reverse
Trendelenburg
possible
* Raise the bed to obtain the possibility of Trendelenburg
Before using this feature, always ensure:
- That the bed extension (optional) is secured
- That there are no objects or children under the bed.
If the sleep surface is not flat and the patient is heavy or the bed has an air
or water mattress, this operation should be carried out by two people.
- Make sure that the bed and accessories (patient helper, I.V. pole, etc.)
cannot strike any fittings (light fittings, service shafts, etc.).
- Make sure that the patient's arms and legs do not jut out from the sleep
surface.
- The Trendelenburg/Reverse Trendelenburg function cannot be used when
transporting a bedside cabinet on the bed. The nursing staff must ensure
that the physiological state of the patient is compatible with these
movements.
4.5
HEAD SECTION
Ensure that there are no obstacles (patient’s limbs, power cables, objects
or accessories) before using this feature.
4.5.1 CPR release
The Head Section CPR
release
is
activated
by
disabling the adjustment
drive in CPR emergency
situations. The release is
cushioned.