7.2.2 PREVENTIVE MAINTENANCE

An annual maintenance visit is recommended to keep the bed fully
operational and to ensure optimal use.

The following points should be given particular attention:

- Brake locking system

- Siderail locking mechanisms

- Drive systems (especially jacks), movement control systems (Head

Section, Leg section and Trendelenburg / Reverse Trendelenburg) and
their ancillary parts

- Bed movement and ancillary part bearings

- Control operation (in particular return to neutral)

- The state of cables and electrical components

- Correct earthing of metal parts and equipotential connectors

- Waterproofing of electrical connections

- Cabling

- Correct condition of the frame and welded assemblies (corrosion and

shocks).

The whole electrical system should be inspected by an approved after-sales
service on average once every three years in order to ensure continued
operation in optimal conditions.

The frequency of the maintenance visits should be determined according
to the condition of the bed and its specific usage, as some components may
need changing after a given period of use (see Maintenance Manual).

Содержание Hill-Rom Evolution LI156E Series

Страница 1: ...User manual Evolution LI156Ex QD1532A Rev002...

Страница 2: ...or otherwise other than for the user s personal use without the prior written permission of Hill Rom PLEASE NOTE Persons depicted in the illustrations may be shown working in non standard working pos...

Страница 3: ...ipment and the transfer of patients to examination wards TABLE OF CONTENTS 0 INTRODUCING YOUR MODEL 3 1 SYMBOLS 5 1 1 GENERAL SYMBOLS 5 1 2 FUNCTION SYMBOLS 6 2 OVERVIEW 8 3 INSTALLING THE BED 9 3 1 F...

Страница 4: ...EDAL AD090A 39 5 9 HAND PENDANT CONTROL SOCKET AD089A 39 5 10 SIDERAIL PROTECTIVE COVER AD113A 39 5 11 OXYGEN CYLINDER HOLDER AC959A AD101A AD102A 40 5 12 ACCESSORY HOLDER 41 6 CLEANING 42 6 1 SAFETY...

Страница 5: ...djustable leg section Mechanical Yes Integral siderails Yes Central brake and steer system Yes Castors 150 Single band castors OPTIONS Trendelenburg Reverse Trendelenburg mandatory for acute care Inte...

Страница 6: ...lder for frame AC962A 3 litre bottle holder for foot end AC963A Syringe driver holder AC967A Monitor stand AC968A Equipotential cable AC991A Triangular section drainage holder AC992A Mattress for bedf...

Страница 7: ...MBOLS 1 SYMBOLS 1 1 GENERAL SYMBOLS Safety instructions DO NOT BIN DO NOT PLACE ANYTHING Please read Follow recycling on this surface carefully procedure Safety earth Equipotential terminal Class II d...

Страница 8: ...djustable Leg section 1 2 2 Brake and steer controls Steer Neutral STOP Brake 1 2 3 Angle control 1 2 3 1 Bed without Bedside Cabinet Transport option Reverse Trendelenburg availability area Head Sect...

Страница 9: ...ion adjustment Knee section adjustment Lock out panels Electrical functions released Electrical functions locked out Variable Height bilateral pedal optional accessory ref AD098A Down Up Hand pendant...

Страница 10: ...l Central brake and steer pedal Electrical function lock out Leg Section control handle Chassis protecting cover Battery 2 accessory holders 150 mm castors Siderail positioning handle Head Section CPR...

Страница 11: ...e ensure that security and conformity requirements are met Before installing the bed for the first time or after bringing the bed and its accessories out of storage Ensure that the bed and its various...

Страница 12: ...nt protection and reduce the risk of falls In the intermediate position the rails can help a fully mobile patient get in and out of bed safely and comfortably Nursing staff must decide which rail posi...

Страница 13: ...ull the rail out from under the bed until fully extended Rotate the rail upwards into the required position intermediate or fully raised Ensure that the siderail has clicked into position 4 1 2 Loweri...

Страница 14: ...1 Patient controls Patients can use the controls located on the inner side of the Head Section rails to adjust Head Section inclination and Auto contour if available see sect 4 3 Knee Section Inclina...

Страница 15: ...ard movements and on the siderail controls 4 2 2 Bilateral variable height foot pedal optional This pedal is located centrally under the bed chassis Nursing staff can use this pedal to adjust bed heig...

Страница 16: ...ndant control Controls released Controls locked out To lock out an electrical function Switch the relevant control switch to lock To release an electrical function Switch the relevant control switch t...

Страница 17: ...pole etc cannot strike any fittings light fittings service shafts etc Make sure that the patient s arms and legs do not jut out from the sleep surface 4 4 2 Variable Height bilateral pedal option Thi...

Страница 18: ...nlock the safety catch by pulling the lever located behind the footboard handle Tilt the sleep surface in the position required Release the safety catch lever Hold the bed by the footboard rail and ti...

Страница 19: ...ildren under the bed If the sleep surface is not flat and the patient is heavy or the bed has an air or water mattress this operation should be carried out by two people Make sure that the bed and acc...

Страница 20: ...bled after the instant release system has been activated whatever the position of the Head Section The CPR emergency release function is unavailable when transporting the bedside cabinet on the bed Th...

Страница 21: ...ture can prevent the patient from sliding down in the bed The Auto contour feature is only available when both Head Section and Knee section adjustment are enabled When the Head Section is locked out...

Страница 22: ...TION 4 6 1 Electrical Knee section control or or Knee section inclination can be adjusted by using the patient or nurse controls or by the hand pendant controls When the Knee section is raised the Leg...

Страница 23: ...raise the Leg section Use the pull out handle on the end of the Leg section to raise it to the required position Carry on lifting until the Leg section locks into position To lower the Leg section Li...

Страница 24: ...t hand wheel is used to steer wheel swivel is disabled The bed can be moved in a straight line 4 7 2 Moving the bed Before moving the bed Ensure that siderails are raised to prevent falls if there is...

Страница 25: ...c To move the bed Grip the footboard or frame with both hands Use the brake and steer pedal to unlock the wheels Push the bed steering with the footboard or frame When moving in a straight line positi...

Страница 26: ...n the bed the power cord must be rolled up on the headboard holder before replacing the headboard 4 10 HEADBOARD AND FOOTBOARD REMOVAL No special tools or devices are required to remove and place the...

Страница 27: ...re below Ensure that the power cord is not wound on one of the hooks 4 11 FRAME The four sections of the bed can be removed for easy cleaning Handle as shown in the picture and ensure that they are se...

Страница 28: ...sed by the straps the electrical functions must be locked out and the bed must be fitted with either accessory AD113A Siderail protective cover or a system of ankle restraints must be used 4 13 FOOT E...

Страница 29: ...g removed when used as a linen holder 4 15 BATTERY optional The power unit automatically charges the battery Maximum life is approximately three months when not connected to mains supply An alarm soun...

Страница 30: ...e the bedside cabinet doors and drawer and put the eating tray on the bed in its folded position Release the brake move the bed away from the wall to free the head end section in order to fit the beds...

Страница 31: ...dside cabinet then apply the brakes Connect the mains power cord Remove the headboard Unlock the electric function of the bed if necessary Operate the Variable Height lower function so that the bedsid...

Страница 32: ...that the Egress bar is over the bed Correct position Incorrect position 5 1 2 Adjustable patient helper type AD081C Safe Working Load 75 kg 1 The adjustable patient helper three positions can be fitt...

Страница 33: ...patient helper Place the patient helper in the required position Turn the knob the other way to lock in position Turn the patient helper slightly until it locks into place Pull patient helper to ensu...

Страница 34: ...justed according to the patient preference To adjust the height of the patient handle push on the locking mechanism button adjust the height and release the button Check the locking mechanism by pulli...

Страница 35: ...shown in the following illustrations 5 2 1 Using the I V Pole AD082A To lengthen the I V pole Raise the upper part of the pole to the desired height To shorten the I V pole Hold the upper part of the...

Страница 36: ...to the desired height Tighten the knob To lower the I V pole Loosen the knob Hold the upper part of the pole firmly and lift slightly while pressing upward on the plastic sleeve with the same hand Low...

Страница 37: ...hten PLEASE NOTE Introduce both ends simultaneously to avoid the frame getting stuck in the sockets The monitor stand is fitted in the same way ref AC967A 5 4 BED EXTENSION PANEL AC992A The panel fits...

Страница 38: ...Load 15 kg 1 The tray can be used for small objects and clothes The tray fits over the chassis bars under the headboard and up against the chassis cover Fitting the tray Introduce the tray sideways be...

Страница 39: ...for use by patients who cannot reach the patient siderail controls The hand pendant control can be used in the same way as the patient siderail controls see section 4 2 1 1 The hand pendant control u...

Страница 40: ...ent is able to use the siderail controls the hand pendant control should be disconnected see section 5 7 3 5 7 4 Storage When not in use the control must be kept in its storage space under the sleep s...

Страница 41: ...leaflet for instructions on how to install See section 4 4 2 for instructions for use The accessory must be fitted by an authorised technician 5 9 HAND PENDANT CONTROL SOCKET AD089A This accessory is...

Страница 42: ...ing staff Check that cylinder is correctly positioned at the base of the Cylinder holder Never use a different oxygen cylinder model from the model that is specified above danger of dropping or interf...

Страница 43: ...ic to certain bed models and cannot be used with other models Due to the diversity of accessories options it is always advisable to ensure that they are mutually compatible load risk of bumps or colli...

Страница 44: ...ailure to implement one or more of these recommendations may lead to damage or deterioration preventing use of the bed and rendering the warranty void 6 2 GENERAL ADVICE NOTE 1 Staining disinfectant p...

Страница 45: ...One pair of household gloves Detergent disinfectant solution diluted according to hospital guidelines and taking into account the recommendations given above or a disinfecting spray Use a standardised...

Страница 46: ...3 4 DECONTAMINATION RECORD A decontamination record should be kept for each bed mentioning Date month ward and room number bed reference number Cleaning frequency materials and products used ANY INFEC...

Страница 47: ...surface to adjust the bed to the required height when the patient is undergoing treatment 7 1 3 Trendelenburg Reverse Trendelenburg optional This feature is mandatory in intensive care units and must...

Страница 48: ...t intended to receive immobilisation equipment e g straps Always keep the bed in its lowest position when the patient is left unattended Given the many different types of mattresses available it is no...

Страница 49: ...onditions and with what type of bed and mattress and what immobilisation or special measures must be used refer to the specific recommendations for half length siderails that follow 2 In which situati...

Страница 50: ...85 88 cm suitable for the sleep surface dimensions If the bed is fitted with an electrically powered dynamic mattress the power cord must be stored so as to prevent it from being cut by the moving pa...

Страница 51: ...that also comply with the electromagnetic standards in force Some devices particularly older ones may however experience interference or may themselves interfere with the working of this product The...

Страница 52: ...PRECAUTIONS Only duly qualified and authorised staff are authorised to carry out maintenance Before any maintenance or repair work Lock out all electrical functions Disconnect the bed Ensure that the...

Страница 53: ...cillary part bearings Control operation in particular return to neutral The state of cables and electrical components Correct earthing of metal parts and equipotential connectors Waterproofing of elec...

Страница 54: ...Kg with the Bedside Cabinet Transport option Connection to a non standard power supply Connection of other electrical appliances to the bed Use of accessories and equipment other than those specified...

Страница 55: ...ring storage the bed must Be in the low position Have all functions locked out Be covered and all moving parts secured Be damp proofed Be kept at a temperature of between 0 and 40 C Beds must never be...

Страница 56: ...9 IPX4 Electric shock protection Class I Intermittent operation 6 min h Safe Working Load 1 Basic version 170kg Bedside Cabinet Transport option 220kg Maximum lifting load 250kg Bed weight no mattress...

Страница 57: ...astors 1 Double band 150 castors 1 Minimum height low position mm 452 448 440 Maximum height high position mm 792 788 780 for a bed equipped with the Bedside Cabinet Transport option add 10mm Head Sec...

Страница 58: ...id in part or in total in the event of Unauthorised interference with or incorrect maintenance of Actuators Electrical drives and components Mechanical systems Abnormal use Use of parts and accessorie...

Страница 59: ...8 01 01 Complies with French standards NF S 90 312 1984 EN 60601 1 1996 EN 60601 1 2 2001 EN 60601 2 38 1999 IEC 60601 2 38 Amendment 1 1999 Complies with essential requirements of CE directive 93 42...

Страница 60: ...Evolution Electric Bed LI156Ex 58 QD1532A 2 USER MANUAL WARRANTY AND AFTER SALES SERVICE Notes...

Отзывы: