Evolution™ Electric Bed LI156Ex
18
QD1532A(2)
USER MANUAL
IN
S
T
RUCT
IO
NS
F
O
R
U
S
E
Use the two handles located under the Head
Section behind each symbol to operate the CPR
release.
To lower the Head Section:
- Release the Head Section by pulling the CPR handle with one hand
- Control the descent of the Head Section using your other hand as
necessary
NOTE: Due to the dampened motion and depending on the weight or
position of the patient, it may be necessary to initially push down on
the head section.
- Release the handle when the Head Section is horizontal.
The Head Section power drive is automatically re-enabled after the instant
release system has been activated, whatever the position of the Head
Section.
The "CPR emergency release" function is unavailable when transporting
the bedside cabinet on the bed. The nursing staff must ensure that the
physiological state of the patient is compatible with these movements.
4.5.2 Head Section control
or
or
or
Head Section adjustment can be activated by the patient / nurse controls or
by the hand pendant controls.
The "Up" button on the Head
Section
adjustment
control
activates the Auto-contour feature.
The
Knee
section
adjustment
control should be locked out using
the lock-out panel if only the Head
Section height adjustment feature
is required.
NOTE: When using the Bedside Cabinet Transport function, lock out the
Head Section control. See precautions, section 4.17.